Skip to main content

Table 1 The differences of esophageal motility measurements between the 22-channel WPM and 36-sensor SSM during wet swallows (n = 12)

From: A comparative study of 22-channel water-perfusion system and solid-state system with 36-sensors in esophageal manometery

 

22-channel WPM

36-sensor SSM

Z value

P

LES proximal margin (cm)

41.5(38-43)

42(39-43)

-2.11

0.035

LES length (cm)

4(4-4)

3.5(3-4)

0.00

1.000

LES rest pressure (mmHg)

13.5(10-21)

16.5(14-22.5)

-1.69

0.091

LES relax ratio (%)

97.5(92-98.5)

84(77-88.5)

-2.20

0.028

Distal esophageal body contraction intensity (mmHg)

68.5(40-85.25)

80(50-126.5)

-1.88

0.060

Distal esophageal body contraction duration (s)

3.85(2.9-4.2)

3.7(3.4-4.525)

-1.18

0.238

Proximal esophageal body contraction intensity (mmHg)

39(22-54)

51.5(40-62.75)

-1.96

0.050

Proximal esophageal body contraction duration (s)

3.05(2.6-3.275)

3.2(2.5-3.325)

-0.67

0.505

UES proximal margin (cm)

17(16-18)

16.5(16-17.25)

-0.81

0.417

UES length (cm)

4(4-5)

4(3-5)

-0.52

0.603

UES rest pressure (mmHg)

56.5(43-82.5)

77.5(60-113)

-1.88

0.060

UES relax ratio (%)

100(100-100)

100(97-100)

-0.76

0.446