Skip to main content

Table 2 Evaluation of the number of false-positive findings for each visibility score

From: Capsule endoscopy with flexible spectral imaging color enhancement reduces the bile pigment effect and improves the detectability of small bowel lesions

 

No. of Segments

No. of false positive findings/segment

Conventional

Setting 1

Setting 2

Setting 3

Visibility score

     

 excellent

7

0.1 ± 0.4

0.1 ± 0.4

0.3 ± 0.5

0.4 ± 0.5

 good

13

0.3 ± 0.1

0.2 ± 0.6

0.2 ± 0.4

0.5 ± 0.7

 fair

20

0.6 ± 0.8

0.9 ± 1.0

1.2 ± 1.1

1.7 ± 1.4*

 poor

8

0.9 ± 0.6

1.6 ± 0.9

2.9 ± 1.9†

2.6 ± 1.8†

  1. NOTE: Each of the 12 CE videos was divided into four segments of equal length, according to the small bowel transit time, and landmarks were placed to identify the last 10 minutes of every quartile. The CE experts evaluated the visibility of the each segment of 10 minutes' duration using a 4-point qualitative index (excellent, good, fair and poor).
  2. Values are mean ± SD.
  3. *Comparison of conventional CE with FICE setting 3 by the Mann–Whitney U test (P = 0.018).
  4. †Comparison of conventional CE with FICE settings 2 and 3 by the Mann–Whitney U test (P = 0.046 and P = 0.024).